Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual
Thermo Fisher Scientific pH 5+, pH 6+, and Ion 6+ are microprocessor-based handheld meters with a large custom LCD. The pH 5+ measures pH and temperature in °C; the pH 6+ and Ion 6+ also measure mV (ORP) and temperature, and the Ion 6+ measures direct ion concentration of various mono- and divalent ions.
This manual explains keypad functions, LCD annunciators, battery replacement, electrode and temperature sensor connections, pH electrode conditioning, switching on, calibration, measurement, electrode care, troubleshooting, specifications, accessories, warranty, and return procedures.
Thermo Fisher Scientific pH 5+ specifications
| Ion range | 0.01 to 1999 ppm |
|---|---|
| Ion resolution | 0.01 ppm for 0.01 to 0.99 ppm; 0.1 ppm for 1.0 to 199.9 ppm; 1 ppm for 200 to 1999 ppm |
| Ion accuracy | +/- 1% of reading |
| Ion calibration points | 2 to 3 points (minimum 2 pts) |
| pH range | 0.00 to 14.00 pH |
| pH slope range | 80 to 120% |
| pH calibration points | 1 to 3 points (push-button) |
| Buffer options | pH 4.01, 7.00, 10.01 (USA); pH 4.01, 6.86, 9.18 (NIST); pH 4.10, 6.97 (Pb) |
| Temperature range | 0.0 to 100.0 °C |
| Temperature compensation | Automatic / Manual (0 to 100 °C) |
| Millivolt range | -1000 to +1000 mV |
| Millivolt range | -500 to 500 mV |
| Millivolt resolution | 0.1 mV for -200 to 200 mV; 1 mV for 200 to 500 mV |
| Millivolt accuracy | +/- 0.2 and 2 mV respectively |
| Auto-buffer recognition | Yes |
| Hold function | HO |
| Auto shut off | After 17 minutes |
| Low battery indication | LO |
| Display | Single Custom LCD |
| Operating temperature | 0 to 50 °C |
| Power requirements | 4 x AAA alkaline batteries |
| Battery life | 500 hours |
| Meter dimensions/weight | 15.7 x 8.5 x 4.2 cm / 255 g |
Source: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual
All Thermo Fisher Scientific pH 5+ documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What does the LO annunciator indicate?
It indicates a low battery condition.
From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual
How should a pH electrode be conditioned before use?
Soak it in pH 4 buffer solution for at least 1 hour, then rinse before use.
From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual
What are the measurement modes for the pH 5+ and pH 6+ meters?
The pH 5+ measures pH and temperature (oC). The pH 6+ measures pH, mV (ORP), and temperature.
From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual
How do I calibrate the meter?
Calibrate using pH buffer standards (USA, NIST, or Pb groups) and follow the step-by-step procedure in Section 3.1. Calibration values are automatically saved.