Key facts
- Brand
- Sponsler
- Model
- SP712-2
- Category
- flow transmitter
- Industry
- Industrial
- Documents
- 2
- Languages
- EN
- Last added
- 2026-11-09
Start here

Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual
This is the installation and operation manual (DOC# MN-712-C, Rev. C) for the Sponsler SP712-2, a 2-wire loop powered analog transmitter that linearly converts a frequency input to an equivalent 4-20mA current output. Used with a turbine flowmeter, it provides a current representation proportional to flow, and current-format data transmission gives noise immunity and long distance transmission capability. A full-scale frequency range of 100 Hz - 10 KHz is selectable via switch S1.
The manual lists the transmitter specifications, including operating and storage temperatures, input voltage limits, signal input, analog output, response time, linearity, tempco, enclosure ratings and weight. It then gives a bench test calibration procedure using a power supply, digital multimeter, frequency generator and frequency counter, with zero (R25) and span (R19) adjustments, sensitivity adjustment (R1) and a linearity check formula, followed by typical loop configurations, a terminal location and outline drawing, and the note that lamp D1 illuminates while test button S2 is depressed if both signal and power are present.
Sponsler SP712-2 specifications
| Operating temperature | -40 to 85 °C |
|---|---|
| Storage temperature | -65 to 125 °C |
| Minimum input voltage | 7 V + (20mA x RL) |
| Maximum input voltage | 28 V + (4mA x RL) |
| Signal input frequency | 0 - 10 KHz |
| Signal input amplitude | 50 mV - 35 V sine or square wave |
| Input impedance | 50K |
| Analog output | 4mA @ 0 Hz, 20mA @ desired full scale frequency |
| Full scale frequency range | 100 Hz - 10 KHz selectable via S1 |
| Response time | 95% of change in 1 second |
| Linearity | 0.3% F/S |
| Tempco | < 2% of reading over entire temperature range |
| Enclosure | FM Approved, CSA Certified Class I Groups B, C, D; Class II Groups E, F, G |
| Weight | 1.7 lbs |
Source: Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual
All Sponsler SP712-2 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Mounting, wiring and first power-up. Read it before the hardware goes in.
Questions answered in these manuals
How many wires does the SP712-2 transmitter need for installation?
Only two wires, because it is a true 2-wire transmitter: input power and signal output use the same wires.
From: Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual
What is the selectable full scale frequency range?
A full scale frequency range of 100 Hz to 10 KHz is selectable via switch S1.
From: Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual
How do you calibrate the SP712-2 transmitter?
Use the bench‑test calibration procedure: connect power, DMM, and frequency generator as shown, set zero to 4 mA, adjust span to 20 mA, and verify linearity at selected frequencies.
From: Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual
What is the operating temperature range of the SP712-2?
The transmitter operates from –40 °C to 85 °C and can be stored from –65 °C to 125 °C.
From: Sponsler SP712-2 Loop Powered 4-20mA Transmitter Installation & Operation Manual