The Sponsler SP712-2 manual provides installation, operation, specifications, and calibration instructions for the loop‑powered 4‑20 mA transmitter.
This manual describes the SP712-2 loop‑powered 4‑20 mA transmitter, a 2‑wire device that converts a frequency input to a proportional current output. It includes key features such as selectable full‑scale frequency range, adjustable sensitivity, and a compact 1.7 lb enclosure suitable for direct mounting on flowmeters.
The document contains detailed specifications, wiring diagrams, typical loop configurations, and a step‑by‑step bench‑test calibration procedure. Users can follow the calibration steps to set zero, span, and sensitivity, and verify linearity across the selectable frequency range.
| Operating Temperature | -40 to 85 °C |
|---|---|
| Storage Temperature | -65 to 125 °C |
| Input Voltage | Min 7 V + (20 mA×RL), Max 28 V + (4 mA×RL) |
| Signal Input Frequency | 0‑10 kHz |
| Signal Input Amplitude | 50 mV – 35 V sine or square wave |
| Impedance | 50 kΩ |
| Analog Output Current | 4 mA @ 0 Hz, 20 mA @ full‑scale frequency |
| Full‑Scale Frequency Range | 100 Hz‑10 kHz (selectable) |
| Response Time | 95 % of change in 1 s |
| Linearity | ±0.3 % |
| Weight | 1.7 lb |
Use the bench‑test calibration procedure: connect power, DMM, and frequency generator as shown, set zero to 4 mA, adjust span to 20 mA, and verify linearity at selected frequencies.
The transmitter operates from –40 °C to 85 °C and can be stored from –65 °C to 125 °C.
It needs a minimum of 7 V plus the loop voltage (20 mA×RL) and a maximum of 28 V plus the loop voltage (4 mA×RL).