ResMed HumidAire 2i Manuals

ManualLib has 3 documents for the ResMed HumidAire 2i (Health Care): 2 user manuals, 1 service manual. Every file can be read online or downloaded as a PDF for free.

User Manual 2Service Manual 1

Key facts

Brand
ResMed
Model
HumidAire 2i
Category
Health Care
Industry
Health
Documents
3
Languages
JA · EN
Last added
2011-11-16

Start here

ResMed HumidAire 2i Disinfection and Sterilization Guide

ResMed HumidAire 2i Disinfection and Sterilization Guide

User Manual 11 pages · 555 KB · JA

This guide provides the recommended procedures for cleaning, disinfecting and sterilizing the HumidAire 2i chamber in clinical settings. It covers four options ranging from manual cleaning with high‑heat disinfection to automatic cleaning and chemical or steam disinfection.

The document also includes step‑by‑step disassembly and reassembly instructions, storage recommendations, and troubleshooting tips for water leaks after reassembly. This manual is written in Japanese.

Read online

ResMed HumidAire 2i specifications

Storage temperature range-20 °C to 60 °C
High‑heat disinfection temperature 1132 °C, hold 3 min
High‑heat disinfection temperature 2121 °C, hold 30 min
Cleaning solution concentration (Option 1 & 2)Neodisher Med1Zym 2 % dilution

Source: ResMed HumidAire 2i Disinfection and Sterilization Guide

All ResMed HumidAire 2i documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Service Manual 1

Disassembly, diagnostics and part numbers, for repair.

Questions answered in these manuals

What are the recommended temperature and time for high‑heat disinfection?

Use either 132 °C for a 3‑minute hold or 121 °C for a 30‑minute hold, depending on the equipment’s capability.

From: ResMed HumidAire 2i Disinfection and Sterilization Guide

How should the chamber be stored after cleaning?

Store it in a dry, dust‑free place away from direct sunlight, at temperatures between –20 °C and 60 °C.

From: ResMed HumidAire 2i Disinfection and Sterilization Guide

Which cleaning solution is recommended for the HumidAire 2i?

The recommended cleaning solution is Neodisher MediZym at a 2 % concentration.

From: ResMed HumidAire 2i Sterilization Manual

What cleaning solution is recommended for the HumidAire 2i?

The guide recommends using Neodisher MediZym at a 2% concentration.

From: ResMed HumidAire 2i Sterilization Guide

What alternative cleaning agent can be used with the HumidAire 2i?

Alconox at a 1% concentration is listed as an alternative cleaning agent.

From: ResMed HumidAire 2i Sterilization Guide