This document is the disinfection and sterilization guide for ResMed HumidAire 2i™, which is intended for use in sleep labs, clinics, or hospitals. The document provides instructions for cleaning, disinfecting, and sterilizing the water tank.
This manual provides detailed instructions for sterilizing and cleaning the HumidAire 2i water tank in clinical environments. It covers four disinfection methods: manual cleaning with high-temperature disinfection, automatic cleaning with high-temperature disinfection, manual cleaning with chemical disinfection, and manual cleaning with steam sterilization. Each method includes steps for disassembly, pre-cleaning, cleaning, disinfection, drying, inspection, and reassembly. The manual also specifies storage conditions and safety precautions for handling the device. This manual is written in Chinese.
| Storage Temperature | –20°C to 60°C |
|---|---|
| Sterilization Methods | Manual Cleaning with High-Temperature Disinfection, Automatic Cleaning with High-Temperature Disinfection, Manual Cleaning with Chemical Disinfection, Manual Cleaning with Steam Sterilization |
Four methods are recommended: manual cleaning with high-temperature disinfection, automatic cleaning with high-temperature disinfection, manual cleaning with chemical disinfection, and manual cleaning with steam sterilization.
The storage temperature should be between -20°C and 60°C.
If the water tank shows any obvious damage such as cracks, fractures, or breaks, it should be discarded and replaced immediately.