Eppendorf CryoCube F740hi, F740hiw User Manual
This manual provides detailed instructions for installing, operating, and maintaining the Eppendorf CryoCube F740hi and F740hiw ultra-low temperature freezers. It covers safety procedures, product features, setup, user interface navigation, alarm and event management, maintenance tasks, and troubleshooting. The document also includes technical specifications, accessory information, and guidelines for transport and disposal.
Read onlineEppendorf CryoCube F740h specifications
| Average energy savings at -70 °C versus -80 °C | about 30% |
|---|---|
| Test basis for energy savings data | Eppendorf-external tests with three empty units (230 V) of each type in parallel at 20 °C room temperature |
| Pull-down time to -80 °C | approximately 4 hours |
| Recommended clearance from back wall | 150 mm / 6 inch |
| Recommended clearance from lateral objects | at least 150 mm / 6 inch |
| Recommended room temperature at freezer location | +20 °C |
Source: Eppendorf CryoCube F740h ULT Freezer Energy Savings Leaflet
All Eppendorf CryoCube F740h documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How much energy can be saved by changing the freezer set point from -80 °C to -70 °C?
On average about 30% energy savings with CryoCube models, based on Eppendorf-external tests with three empty 230 V units of each type at 20 °C room temperature.
From: Eppendorf CryoCube F740h ULT Freezer Energy Savings Leaflet
How long does a CryoCube freezer take to pull down to -80 °C?
It can pull down to -80 °C within approximately 4 hours, so it does not need to run constantly as a back-up freezer.
From: Eppendorf CryoCube F740h ULT Freezer Energy Savings Leaflet
How do I install the CryoCube F740hi/F740hiw?
Install by selecting a location, unpacking, checking delivery, transporting to location, connecting to voltage and cooling water supply, and turning on the device.
What interfaces does the CryoCube F740hi/F740hiw have?
It has remote alarm interface, RS-485 interface, and Ethernet interface for connecting to external systems.
How do I register the CryoCube F740hi/F740hiw?
Register the device via the VisioNize onboard system or by visiting www.eppendorf.com/myeppendorf.
What interfaces does the CryoCube F740hi/F740hiw have?
It has RS-485, Ethernet, and remote alarm interfaces for external system connectivity.