Eppendorf CryoCube F740h ULT Freezer Energy Savings Leaflet

Eppendorf leaflet on saving energy with CryoCube ULT freezers by moving the set point from -80 °C to -70 °C, plus installation, maintenance and sample tips.


Eppendorf CryoCube F740h ULT Freezer Energy Savings Leaflet - cover page
Brand
eppendorf
Category
Professional medical equipment
Document type
Product Leaflet
Language
English
Pages
1
File format
PDF
File size
841 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
2F180BF60BEDE1435B6E9B6C83D25684

About this manual

This one-page Eppendorf leaflet explains how to save energy with Eppendorf ULT freezers, using the CryoCube F740h, F740hi, F570h, F570n, F440h and F440n as examples. It states that changing the set point from -80 °C to -70 °C gives average energy savings of about 30%, based on Eppendorf-external tests with three empty 230 V units of each type in parallel at 20 °C room temperature. It notes that most samples should not be impacted by the change.

The leaflet also gives practical advice: keep 150 mm/6 inch clearance from the back wall and lateral objects, keep the freezer top free of material, and check the room temperature (+20 °C recommended). Further sections cover avoiding constantly running back-up freezers because a CryoCube freezer can pull down to -80 °C in about 4 hours, training users, regular maintenance such as cleaning the air filter and annual defrosting, running a yearly Freezer-Sample Day, and using standardized boxes, racks and sample management software to maximize storage space.

What's inside

  • -70 °C instead of -80 °C: -30% Energy Savings
  • Installation
  • Back-up Freezer
  • Training
  • Maintenance and Service
  • Freezer-Sample Day
  • Sample Management

Specifications

Average energy savings at -70 °C versus -80 °Cabout 30%
Test basis for energy savings dataEppendorf-external tests with three empty units (230 V) of each type in parallel at 20 °C room temperature
Pull-down time to -80 °Capproximately 4 hours
Recommended clearance from back wall150 mm / 6 inch
Recommended clearance from lateral objectsat least 150 mm / 6 inch
Recommended room temperature at freezer location+20 °C

Frequently asked questions

How much energy can be saved by changing the freezer set point from -80 °C to -70 °C?

On average about 30% energy savings with CryoCube models, based on Eppendorf-external tests with three empty 230 V units of each type at 20 °C room temperature.

How long does a CryoCube freezer take to pull down to -80 °C?

It can pull down to -80 °C within approximately 4 hours, so it does not need to run constantly as a back-up freezer.

How much space should be kept around the freezer during installation?

Keep 150 mm/6 inch from the back wall and at least 150 mm/6 inch from lateral objects, and keep the top of the freezer free of packaging or supply boxes.

Which maintenance tasks improve energy consumption?

Clean the air filter regularly, manually remove snow, frost and ice inside the compartments once a week, keep door gaskets ice-free, and defrost the freezer once a year with basic cleaning.

Download PDF (841 KB)

More eppendorf Manuals


View all eppendorf manuals →