Eppendorf leaflet on saving energy with CryoCube ULT freezers by moving the set point from -80 °C to -70 °C, plus installation, maintenance and sample tips.
This one-page Eppendorf leaflet explains how to save energy with Eppendorf ULT freezers, using the CryoCube F740h, F740hi, F570h, F570n, F440h and F440n as examples. It states that changing the set point from -80 °C to -70 °C gives average energy savings of about 30%, based on Eppendorf-external tests with three empty 230 V units of each type in parallel at 20 °C room temperature. It notes that most samples should not be impacted by the change.
The leaflet also gives practical advice: keep 150 mm/6 inch clearance from the back wall and lateral objects, keep the freezer top free of material, and check the room temperature (+20 °C recommended). Further sections cover avoiding constantly running back-up freezers because a CryoCube freezer can pull down to -80 °C in about 4 hours, training users, regular maintenance such as cleaning the air filter and annual defrosting, running a yearly Freezer-Sample Day, and using standardized boxes, racks and sample management software to maximize storage space.
| Average energy savings at -70 °C versus -80 °C | about 30% |
|---|---|
| Test basis for energy savings data | Eppendorf-external tests with three empty units (230 V) of each type in parallel at 20 °C room temperature |
| Pull-down time to -80 °C | approximately 4 hours |
| Recommended clearance from back wall | 150 mm / 6 inch |
| Recommended clearance from lateral objects | at least 150 mm / 6 inch |
| Recommended room temperature at freezer location | +20 °C |
On average about 30% energy savings with CryoCube models, based on Eppendorf-external tests with three empty 230 V units of each type at 20 °C room temperature.
It can pull down to -80 °C within approximately 4 hours, so it does not need to run constantly as a back-up freezer.
Keep 150 mm/6 inch from the back wall and at least 150 mm/6 inch from lateral objects, and keep the top of the freezer free of packaging or supply boxes.
Clean the air filter regularly, manually remove snow, frost and ice inside the compartments once a week, keep door gaskets ice-free, and defrost the freezer once a year with basic cleaning.