User manual for the YTC YTC2221 digital dual-clamp phase voltmeter, covering operation, specifications, and safety instructions for measuring voltage…
The YTC YTC2221 Digital Dual-Clamp Phase Voltmeter is a high-precision, portable handheld instrument designed for on-site measurement of voltage, current, and phase. It features dual-channel input and allows for the measurement of U-U, I-I, and U-I phase angles without breaking the circuit. The device includes a high-contrast LCD with an adjustable viewing angle of approximately 70 degrees. This manual is written in Chinese.
| Operating Temperature (Reference) | (23±5)℃ |
|---|---|
| Operating Humidity (Reference) | (45~75)% RH |
| Signal Frequency (Reference) | (50±0.2)Hz |
| Voltage Input Impedance | 2MΩ |
| Phase Measurement Error Limit | ±2° |
| Display Digits | 3 1/2 |
| Sampling Rate | 3 times/second |
| Power Source | Single 9V battery |
| Weight (Body) | 280g |
| Weight (Clamps) | 2×200g |
The basic error limit for phase measurement is ±2°.
Press the lock button on top of the display screen and flip the screen to the desired angle.
The LCD screen will show a low battery indicator symbol.
The maximum voltage measurement limit is 500V.