WTW photoLab S6 Photometer User Manual

WTW photoLab S6 photometer operating manual covering setup, concentration and extinction/transmission measurements, documentation, methods, quality…


WTW photoLab S6 Photometer User Manual - cover page
Brand
WTW
Category
Analyzers
Document type
User Manual
Language
German
Pages
60
File format
PDF
File size
1.1 MB
Published
13 November, 2026
Updated
13 November, 2026
MD5 checksum
13DDD8244A6F47A40987176D63F7584C

About this manual

The WTW photoLab S6 is a filter photometer for analyzing constituents in water and aqueous solutions using round or rectangular (OS) cuvettes. It operates at ambient temperatures from +5 °C to +40 °C and performs a self-check of memory, optics, electronic measurement acquisition, barcode recognition and cuvette recognition.

This operating manual describes safety information, commissioning, concentration and extinction/transmission measurements, documentation, method parameters, analytical quality assurance, correction functions, zero adjustment, device setup, method data updates, the RS232 interface, maintenance, cleaning, disposal, technical data and troubleshooting. This manual is written in German.

What's inside

  • Overview
  • Safety
  • Commissioning
  • Concentration measurement
  • Extinction/transmission measurement
  • Documentation
  • Method parameters
  • Analytical quality assurance (AQS)
  • Correction functions
  • Technical data

Specifications

Optical measuring principleFilter photometer with reference-beam absorption measurement; simultaneous capture of all wavelengths
Light sourceTungsten-halogen lamp, pre-adjusted
Receiver6-fold photodiode array
Optical filters340 nm, 445 nm, 525 nm, 550 nm, 605 nm, 690 nm
Photometric reproducibility0.001 E at 1.000 E
Photometric resolution0.001 E
Warm-up timeNone
Measurement timeApprox. 2 s
Measurement modesConcentration (method-dependent, display format adjustable), extinction
Extinction range-0.300 E to 3.200 E
Transmission range0.1 % to 1000 %
AdjustmentPermanently stored
Drift correctionAutomatic at every self-check
New methodsVia internet
Barcode recognitionAutomatic method selection; automatic reagent lot recognition
Self-checkTest of memory, optics, electronic measurement acquisition, barcode recognition, cuvette recognition
Automatic calibrationOptics, electronic measurement acquisition, barcode recognition
Time/dateReal-time clock in device
DimensionsH: 140 mm, T: 270 mm, B: 260 mm
WeightApprox. 2.3 kg (battery version: 2.8 kg)
Device safetyEN 61010, IEC 1010
Protection classEN 61010-1/class 3
Power supplyFRIWO FW 7555O/15 (Friwo Part. No. 1822367) or RiHuiDa RHD20W150100; input 100-240 V AC, 50-60 Hz, 400 mA; output 15 V DC, 1 A
Power consumption in mains operationMax. 1300 mA
Data-retention battery1 x 3.0 V lithium battery, soldered in
Optional batteryBuilt-in NiMH battery 7.2 V/2500 mAh; approx. 40 h operation at 10 measurements/hour; trickle charge in mains operation; approx. 5 h charging time when discharged; deep-discharge protection
EMCDirective 2004/108/EC, EN 61326-1, EN 61000-3-2, EN 61000-3-3, FCC Class A
Climate class2, VDI/VDE 3540
Storage temperature-25 °C to +65 °C
Operating temperature+5 °C to +40 °C
Permissible relative humidity75 % annual mean; 95 % for 30 days/year; 85 % on remaining days

Frequently asked questions

What is the photometer intended to be used for?

Analysis of constituents in water and aqueous solutions using round cuvettes or rectangular (OS) cuvettes.

What is the operating temperature range?

+5 °C to +40 °C.

How long does a measurement take?

About 2 seconds.

What is the weight of the photometer?

Approximately 2.3 kg, or 2.8 kg for the battery version.

Download PDF (1.1 MB)

More WTW Manuals


View all WTW manuals →