Supplement to the Tosoh TSKgel NH2-100 3 μm column instruction manual: installation, flow rate and pressure drop, guard column and quality standards.
This document is a supplementary instruction manual from Tosoh for the TSKgel NH2-100 DC 3 μm normal-phase HPLC column, supplied in a 4.6 mm I.D. × 5 cm L size. It only describes the points that differ from the standard TSKgel NH2-100 3 μm column instruction manual, which should be read first for the safety notes and general handling instructions.
The supplement covers the column's end fitting with an external male connector for direct connection to another HPLC column, the parts of the column, installation and connection cautions, the recommended flow rate and maximum pressure drop, the matching guard column, and the quality standard table with theoretical plate count and asymmetry factor. This manual is written in Chinese.
| Part number | 0021999 |
|---|---|
| Column dimensions | 4.6 mm I.D. × 5 cm L |
| Recommended flow rate | 1.0 mL/min |
| Maximum pressure drop | 5 MPa |
| Theoretical plates | 6,000 TP/column |
| Asymmetry factor | 0.90 to 1.30 |
| Guard column | TSKgel guardgel NH2-100 3 μm, 3.2 mm I.D. × 1.5 cm L (part no. 0021972) |
| Guard column cartridges per box | 3 |
| Maximum tightening torque when joining two columns | 2.0 N·m (about 30°) |
The recommended flow rate is 1.0 mL/min, with a maximum pressure drop of 5 MPa.
TSKgel guardgel NH2-100 3 μm, 3.2 mm I.D. × 1.5 cm L (part no. 0021972); each box contains three guard column cartridges.
The quality standard table gives 6,000 theoretical plates per column and an asymmetry factor of 0.90 to 1.30.
First connect the column outlet and the other column inlet by hand, then tighten with a suitable wrench and check for solvent leaks; torques above 2.0 N·m may permanently damage the column.