The Tosoh TSKgel Phenyl-5PW and Ether-5PW HIC column user manual provides installation, operation, storage, and safety guidelines.
This manual describes the TSKgel Phenyl-5PW and Ether-5PW hydrophobic interaction chromatography (HIC) columns, including safety precautions, unpacking, installation, and proper flow direction. It explains how to store the columns, prepare samples and eluents, and select appropriate flow rates and temperatures for optimal performance.
Additional sections cover guard column usage, column efficiency, troubleshooting, quality specifications, and warranty information, with detailed tables of dimensions, flow rates, pressure limits, and recommended operating conditions.
| Operating temperature range | 10‑30 °C |
|---|---|
| Organic solvent concentration limit | up to 50 % |
| Storage temperature range | 4 °C to 30 °C |
Store the columns at a constant temperature between 4 °C and 30 °C, avoid freezing, keep them away from direct sunlight and corrosive gases, and seal both ends with end plugs for long‑term storage.
Typical flow rates are 0.5‑1.0 mL/min for the 7.5 mm ID steel columns and 0.5‑0.8 mL/min for the 5 mm ID glass columns, with corresponding pressure limits of up to 2.0 MPa.
The columns are usable over a pH range of 2.0 to 12.0 at room temperature.
Avoid admitting any bubbles into the column; remove all bubbles from the piping before installing the column to prevent performance degradation.