1. This laboratory provides TOSOH 2.2 Plus glycosylated hemoglobin (Hgb A1c) assay method. 2. The Tosoh 2.2 Plus is a fully automated glycosylated hemoglobin analyzer that utilizes nonporous ion-exchange high performance liquid chromatography (HPLC) for separation of Hemoglobin A1c. 3. This separation is performed rapidly and precisely without interference from Schiff base (labile A1c), lipemia or temperature fluctuations. 4. A1c analysis is performed without off-line specimen pretreatment or interference from Schiff base (labile A1c). 5. The analyzer dilutes the whole blood specimen with Hemolysis/Wash solution, then injects a small volume of the treated specimen onto the non-porous column. 6. Separation is achieved by utilizing differences in ionic interactions between the cation exchange group on the column resin surface and the hemoglobin components. 7. The hemoglobin fractions are removed from the column by performing a step-wise elution (gradient) using 3 buffers of different salt (ionic) concentrations. 8. The separated hemoglobin fractions pass through an LED photometer flow cell where the analyzer measures changes in absorbance. 9. The analyzer integrates and reduces the data then calculates the A1c level.
The Tosoh 2.2 Plus is a fully automated glycosylated hemoglobin analyzer that uses nonporous ion-exchange high performance liquid chromatography (HPLC) to separate Hemoglobin A1c. The separation is rapid and precise, without interference from Schiff base (labile A1c), lipemia, or temperature fluctuations. The analyzer dilutes whole blood with hemolysis/wash solution, injects onto a non-porous column, and performs step-wise elution with three buffers of different ionic concentrations. Hemoglobin fractions are detected via an LED photometer flow cell, and a chromatogram is printed for each sample.
The assay is used to measure glycosylated hemoglobin in blood, providing an index of mean blood glucose concentration over the past two months, which is crucial for diabetes management. Each analysis requires three minutes. Two calibrators of known A1c concentration are run once a week to adjust calculation parameters for A1c determination.
| Analysis time | 3 minutes per sample |
|---|---|
| Calibration frequency | Weekly (two calibrators) |
| Separation principle | Nonporous ion-exchange HPLC |
| Sample type | Whole blood |
| Interferences avoided | Schiff base (labile A1c), lipemia, temperature fluctuations |
The Tosoh 2.2 Plus uses nonporous ion-exchange high performance liquid chromatography (HPLC) to separate Hemoglobin A1c.
Each analysis requires 3 minutes.
Two calibrators of known A1c concentration are run once a week to adjust calculation parameters.
The assay avoids interference from Schiff base (labile A1c), lipemia, and temperature fluctuations.