The Simpson 501 Insulation Tester Instruction Manual provides operating instructions, safety precautions, measurement procedures, technical specifications…
This instruction manual covers the Simpson Model 501 Insulation Tester, a handheld instrument for measuring insulation resistance of de-energized devices and systems with nominal voltages up to 1000 V. It also tests ground, protective, and equipotential bonding conductors and includes a 1000 V AC and DC voltage measuring range.
The manual contains safety precautions, battery and fuse replacement, operation of the analog display, procedures for measuring AC/DC voltages and insulation resistance, technical specifications, maintenance instructions, and NIST certification details.
| Insulation test voltage range | Up to 1000 V |
|---|---|
| AC/DC voltage measuring range | 1000 V |
| Frequency range | 40 to 200 Hz |
| Maximum overload capacity | 1200 V |
| Insulation test voltages | 50 V, 100 V, 250 V, 500 V, 1000 V |
| Measuring accuracy | 1.5% |
| Batteries | 6 x 1.5 V D size |
It uses 6 standard 1.5 V D size batteries.
The maximum overload capacity in the voltage measuring range is 1200 V.
The selectable nominal voltages are 50 V, 100 V, 250 V, 500 V, or 1000 V.
It complies with IEC 1010-1/EN 61010-1, IEC 1557-2/EN 61557-2, and IEC 1557-4/EN 61557-4.