This manual provides operating instructions, safety precautions, and maintenance guidelines for the Simpson 260 Series 8P Volt-Ohm-Milliammeter.
The Simpson 260 Series 8P is a rugged, accurate, and compact instrument equipped with a mirrored dial to eliminate parallax. It is designed to measure AC and DC voltage, direct current, resistance, decibels, and output voltage. Key features include overload protection for most ranges, individual dual jacks for high-voltage measurements, and an audio alarm for continuity tests.
| Measurement Functions | AC and DC voltage, direct current, resistance, decibels, Output Voltage |
|---|---|
| Overload Protection Ranges | All ranges except 1V, 10A, 50 µA, 250mV, 500/1000V AC and DC |
| Overload Relay Trigger | Approximately three to four times rated full scale voltage |
| Basic Fuse Rating | 3 AG type, 1 Amp, quick-blow fuse |
| High Energy Fuse Rating | Bussman Type BBS 2A600V or Littlefuse Type BLS 2A 600V |
| Internal Battery 1 | NEDA 13F size D cell (1.5 volts) |
| Internal Battery 2 | NEDA 1604A alkaline battery (9 volts) |
The instrument is equipped with a mirrored dial to eliminate parallax during readings.
Depress the reset button located on the front panel to the right of the output jack after removing the overload.
It may indicate a dangerous condition such as excessive circuit-to-ground voltage; do not assume the relay is faulty.
All ranges except 1V, 10A, 50 µA, 250mV, and the 500/1000V AC and DC ranges.