Roche cobas Cdiff Nucleic Acid Test for cobas Liat System Quality Control Lot Validation Guide

Roche cobas Cdiff Nucleic Acid Test lot validation guide explains the Add Lot procedure for new assay tube lots on the cobas Liat System.


Roche cobas Cdiff Nucleic Acid Test for cobas Liat System Quality Control Lot Validation Guide - cover page
Brand
Roche
Category
Health Care
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
72 KB
Published
23 November, 2026
Updated
23 November, 2026
MD5 checksum
03768E04B1918AEFADF766E814463757

About this manual

This Quality Control Lot Validation Guide covers the Add Lot procedure for the cobas Cdiff Nucleic Acid Test on the cobas Liat System. Before using a new lot of cobas Cdiff assay tubes, the procedure must be performed on the cobas Liat Analyzer to validate the lot at the user's site.

Required materials include the new assay tubes, Package Insert ID Barcode card, cobas Liat Cdiff Positive and Negative Controls, and their barcode cards. The guide provides step-by-step instructions for running controls, scanning barcodes, inserting assay tubes, and transferring lot information to other cobas Liat analyzers using the tools function.

What's inside

  • Quality Control Lot Validation Guide
  • Add Lot procedure
  • Materials needed for Add Lot
  • Assay Menu
  • New Lot
  • Scan Insert ID
  • Scan Negative Control ID
  • After Add Lot workflow

Frequently asked questions

What must be performed before using a new lot of cobas Cdiff assay tubes?

The Add Lot procedure must be performed on the cobas Liat Analyzer to validate the lot.

Which controls are used in the Add Lot procedure?

A Negative Control sample and a Cdiff Positive Control sample are used.

What should be done if a control result is rejected?

Repeat the control run, or contact your local Roche representative if repeated runs fail.

How is lot information transferred to other cobas Liat analyzers?

Use the tools function in the cobas Liat Menu to transfer the lot information.

Download PDF (72 KB)

More Roche Manuals