Roche cobas Cdiff Nucleic Acid Test lot validation guide explains the Add Lot procedure for new assay tube lots on the cobas Liat System.
This Quality Control Lot Validation Guide covers the Add Lot procedure for the cobas Cdiff Nucleic Acid Test on the cobas Liat System. Before using a new lot of cobas Cdiff assay tubes, the procedure must be performed on the cobas Liat Analyzer to validate the lot at the user's site.
Required materials include the new assay tubes, Package Insert ID Barcode card, cobas Liat Cdiff Positive and Negative Controls, and their barcode cards. The guide provides step-by-step instructions for running controls, scanning barcodes, inserting assay tubes, and transferring lot information to other cobas Liat analyzers using the tools function.
The Add Lot procedure must be performed on the cobas Liat Analyzer to validate the lot.
A Negative Control sample and a Cdiff Positive Control sample are used.
Repeat the control run, or contact your local Roche representative if repeated runs fail.
Use the tools function in the cobas Liat Menu to transfer the lot information.