Laboratory procedure manual for measuring iron concentration in frozen serum using Roche Cobas 6000 c501 module, including safety, specimen handling…
This manual details the laboratory procedure for measuring iron in frozen serum using the Roche Cobas 6000 c501 module. It covers test principles, safety precautions, specimen requirements, equipment, reagents, calibration, quality control, and result reporting. The method uses Ferr oZine reagent and is validated for use with NHANES 2019-2020 data. Results are reported in ug/dL, with a reportable range of 5–1063 ug/dL and specific dilution protocols for samples exceeding limits.
The procedure includes handling requirements for serum or lithium heparin plasma, storage conditions, and rejection criteria for hemolyzed or improperly anticoagulated specimens. Calibration is performed daily using Roche Calibrator and must be verified with controls. The analyzer is automated and integrates with a Laboratory Information System (STARLIMS) for data management and reporting.
| Reportable Range | 5–1063 ug/dL |
|---|---|
| Intra-assay %CV (108 ug/dL) | 0.8% |
| Intra-assay %CV (247 ug/dL) | 0.3% |
| Inter-assay %CV (77 ug/dL) | 2.7% |
| Inter-assay %CV (251 ug/dL) | 1.7% |
| Analytical Measurement Range | 5–500 ug/dL |
Results are reported between 5–1063 ug/dL. Values outside this range are automatically diluted or flagged for manual review.
Specimens must be stored at -70°C for long-term stability, or at 4–8°C for up to 3 weeks, or at 20–25°C for up to 7 days.
Only serum or lithium heparin plasma are acceptable. EDTA, citrate, or oxalate will bind iron and interfere with the assay.
Daily two-point calibration using H2O + C.F.A.S. calibrator is required. Reagent lot changes or major repairs require recalibration.