A guide with step‑by‑step troubleshooting and alignment procedures for the Renishaw InVia Reflex MicroRaman system.
The troubleshooting guide outlines general rules for resolving errors on the InVia Reflex MicroRaman, including restarting the software and system when unexpected behavior occurs. It also lists recommended actions that appear in the health‑check dialog and explains the order in which alignment steps should be performed.
Specific sections cover the most common issues such as laser mis‑alignment, silicon reference sample alignment, CCD area mis‑alignment, slit mis‑alignment, and laser brightness problems. Each issue provides a clear fix using the Auto Align functions in the Tools → Autoalign → Align menu, followed by a re‑run of the health check.
Run the Auto Align Silicon Reference Sample first, then run Auto Align Laser, accept the changes, and click OK.
After aligning the silicon reference sample and the laser, run Auto Align CCD Area, accept the changes, and click OK.
Run Auto Align Silicon Reference Sample, then Auto Align Laser, then Auto Align CCD Area, and finally Auto Align Slits, accepting each change before proceeding.