Renishaw InVia Reflex MicroRaman Troubleshooting Guide

A guide with step‑by‑step troubleshooting and alignment procedures for the Renishaw InVia Reflex MicroRaman system.


Renishaw InVia Reflex MicroRaman Troubleshooting Guide - cover page
Brand
Renishaw
Category
Other device
Document type
Service Manual
Language
English
Pages
2
File format
PDF
File size
19 KB
Published
07 October, 2026
Updated
07 October, 2026
MD5 checksum
826F596F2F16FF77FD0207D4E68B544B

About this manual

The troubleshooting guide outlines general rules for resolving errors on the InVia Reflex MicroRaman, including restarting the software and system when unexpected behavior occurs. It also lists recommended actions that appear in the health‑check dialog and explains the order in which alignment steps should be performed.

Specific sections cover the most common issues such as laser mis‑alignment, silicon reference sample alignment, CCD area mis‑alignment, slit mis‑alignment, and laser brightness problems. Each issue provides a clear fix using the Auto Align functions in the Tools → Autoalign → Align menu, followed by a re‑run of the health check.

What's inside

  • General troubleshooting rules
  • Laser alignment procedure
  • Silicon reference sample alignment
  • CCD area alignment
  • Slit alignment
  • Laser brightness troubleshooting

Frequently asked questions

What should I do if the laser is not centered in the crosshairs?

Run the Auto Align Silicon Reference Sample first, then run Auto Align Laser, accept the changes, and click OK.

How can I fix a CCD area mis‑alignment?

After aligning the silicon reference sample and the laser, run Auto Align CCD Area, accept the changes, and click OK.

What steps resolve a slit mis‑alignment?

Run Auto Align Silicon Reference Sample, then Auto Align Laser, then Auto Align CCD Area, and finally Auto Align Slits, accepting each change before proceeding.

Download PDF (19 KB)