User manual for Novus Human ATM ELISA Kit NBP2-69891, for quantitative detection of Human ATM in biological samples using Sandwich-ELISA.
This document provides detailed instructions for using the Novus Human ATM ELISA Kit (NBP2-69891) to quantitatively measure Human ATM protein in serum, plasma, and other biological fluids. The kit uses a Sandwich-ELISA principle with pre-coated antibody plates and includes reagents for sample preparation, incubation, washing, and colorimetric detection. The procedure requires a microplate reader with a 450 nm filter and standard laboratory equipment. Results are calculated by comparing optical density to a standard curve.
The kit is intended for research use only and not for diagnostic or therapeutic purposes. Storage conditions and handling precautions are specified for all reagents, including temperature and light protection. Sample preparation varies by type (serum, plasma, cell lysates, tissue homogenates) and includes centrifugation and dilution steps. Reagent preparation involves diluting concentrated components to working solutions, and the assay procedure includes multiple incubation and wash steps before reading the optical density.
| Sensitivity | 0.10 ng/mL |
|---|---|
| Detection Range | 0.16-10 ng/mL |
| Specificity | Recognizes Human ATM with no significant cross-reactivity |
| Repeatability | Coefficient of variation < 10% |
The detection range is 0.16-10 ng/mL.
The sensitivity is 0.10 ng/mL.
The kit detects Human ATM in serum, plasma, cell lysates, tissue homogenates, and other biological fluids.
The optical density should be measured at 450 nm ± 2 nm.