The Nova Stat Profile pHOx Training Manual provides step‑by‑step instructions for system identification, operation, calibration, reagent replacement…
This training manual guides users of the Nova Stat Profile pHOx analyzer through all aspects of daily operation, from initial system identification and sample handling to detailed maintenance procedures. It includes quick‑reference key facts, checklists, and a structured curriculum for classroom or on‑the‑job training.
Key sections cover running an analysis, calibrating the instrument, using and handling controls, replacing reagents and paper, performing flow tests and flushing, routine and periodic maintenance, sensor upkeep, setup configuration, QC setup, and troubleshooting labs. The appendix provides technical assistance contacts, parts ordering information, and warranty replacement procedures.
| Sample Volume (Micro) | 45 µL whole blood |
|---|---|
| Sample Volume (Normal) | 70 µL whole blood |
| pH Measurement Range | 6.50 – 8.00 pH units |
| PCO₂ Measurement Range | 3.0 – 200 mmHg |
| PO₂ Measurement Range | 0 – 800 mmHg |
| pH Slope Limits | 9.1 – 11.6 |
Only heparinized whole blood samples are acceptable; urine, CSF, cardioplegia solution, or other body fluids must not be used.