Adult application guide for the Nellcor MAX N SpO2 adhesive sensor, covering site selection, wrap placement, adhesive dots and site checks.
This two-page adult application guide explains how to apply the Nellcor MAX N SpO2 adhesive sensor. It directs users to read the sensor's Directions for Use insert for complete instructions, warnings and specifications before use, and to contact the Medtronic product representative with questions.
The guide describes placing the adhesive wrap fabric side down, using an index finger or other fingers as the site with the window next to the cable on the nail side distal to the first joint, wrapping the sensor without stretching the tape so the optical components oppose each other, anchoring the cable with medical-grade tape, and plugging the sensor cable into the patient cable attached to the monitor. It also covers using clear double-sided adhesive dots to restore adhesive quality, checking the sensor site at least every eight hours, and notes the sensor is for adults weighing more than 40 kg.
| Patient weight | Adults weighing more than 40 kg |
|---|---|
| Maximum adhesive dots per window | Three, one on top of another |
| Sensor site check interval | At least every eight hours |
The preferred site is an index finger, though other fingers may be used. The window next to the cable goes on the nail side, distal to the first joint, and not on a joint.
A maximum of three dots may be applied to each window, one on top of another. Removing a sticky dot from a window may damage the sensor components.
Check the sensor site, including skin integrity, at least every eight hours and change the sensor site as appropriate.
Contact your Medtronic product representative at 800-635-5267.