Nellcor MAX N SpO2 Adhesive Sensor Adult Application Guide

Adult application guide for the Nellcor MAX N SpO2 adhesive sensor, covering site selection, wrap placement, adhesive dots and site checks.


Nellcor MAX N SpO2 Adhesive Sensor Adult Application Guide - cover page
Brand
Nellcor
Category
sensor
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
743 KB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
100352F471E7F5D8980ACCB81F6B11AB

About this manual

This two-page adult application guide explains how to apply the Nellcor MAX N SpO2 adhesive sensor. It directs users to read the sensor's Directions for Use insert for complete instructions, warnings and specifications before use, and to contact the Medtronic product representative with questions.

The guide describes placing the adhesive wrap fabric side down, using an index finger or other fingers as the site with the window next to the cable on the nail side distal to the first joint, wrapping the sensor without stretching the tape so the optical components oppose each other, anchoring the cable with medical-grade tape, and plugging the sensor cable into the patient cable attached to the monitor. It also covers using clear double-sided adhesive dots to restore adhesive quality, checking the sensor site at least every eight hours, and notes the sensor is for adults weighing more than 40 kg.

What's inside

  • Sensor application steps
  • Site selection
  • Adhesive dots
  • Cable anchoring and connection
  • Sensor site checks
  • Directions for Use reference

Specifications

Patient weightAdults weighing more than 40 kg
Maximum adhesive dots per windowThree, one on top of another
Sensor site check intervalAt least every eight hours

Frequently asked questions

Which site is preferred for the MAX N sensor?

The preferred site is an index finger, though other fingers may be used. The window next to the cable goes on the nail side, distal to the first joint, and not on a joint.

How many adhesive dots can be applied to each window?

A maximum of three dots may be applied to each window, one on top of another. Removing a sticky dot from a window may damage the sensor components.

How often should the sensor site be checked?

Check the sensor site, including skin integrity, at least every eight hours and change the sensor site as appropriate.

Who should I contact with questions about the sensor?

Contact your Medtronic product representative at 800-635-5267.