Product leaflet for MT pressure probes (models 0638.1345 to 0638.1645) covering operation, handling, calibration, and technical specifications.
This document describes the operation, handling, and calibration of MT differential and absolute pressure probes. It explains that probes work based on the strain gauge principle and must be connected with the higher pressure applied to the '+' inlet for best accuracy. The text advises fixing the probe in its usage position, typically lying flat, before measurement to ensure reproducibility, and warns against vibrations during use.
Technical specifications are provided for four models: 0638.1345 (±100 Pa), 0638.1445 (±10 hPa), 0638.1545 (±100 hPa), and 0638.1645 (2 bar absolute). The document details operating temperatures, zero temperature coefficient drift, accuracy, and maximum overload limits for each model. It also lists maintenance-free care instructions and a 24-month warranty.
| Storage and transport temperature | -20 to +70 °C |
|---|---|
| Model 0638.1345 Measuring range | ± 100 Pa |
| Model 0638.1445 Measuring range | ±10 hPa |
| Model 0638.1545 Measuring range | ± 100 hPa |
| Model 0638.1645 Measuring range | 2 bar abs |
| Warranty period | 24 months |
The higher pressure should be applied to the '+' inlet to achieve the best possible accuracy.
Zero calibration is required for differential pressure probes, but not necessary for the absolute pressure probe (0638.1645).
The operating temperature range is 0 to +50 °C.
The warranty period for the pressure probe is 24 months.