User manual for Mitsubishi Electric ME110SSR multi-measuring instrument covering safety, operation, maintenance, and installation instructions.
This user manual provides essential safety precautions and operational guidelines for the Mitsubishi Electric ME110SSR multi-measuring instrument. It includes instructions for proper installation, wiring, and maintenance to ensure safe and accurate operation. The manual also details environmental and storage conditions, disposal guidelines, and EMC compliance requirements for cabinet installation.
The document covers various models in the ME110SSR series, including ME110SSR-4AP, ME110SSR-4APH, ME110SSR-4A2P, ME110SSR-C, ME110SSR-CH, and ME110SSR-MB, with specific wiring diagrams and terminal specifications for each variant. This manual is written in Chinese.
| Auxiliary Power Supply | 100-240V AC+10-15% (50-60Hz) 10VA, 100V DC+40-25% 6W |
|---|---|
| Operating Temperature | -5 to 50°C |
| Humidity | 30~85%RH, non condensing |
| Altitude | 1000m or less |
| Pollution Degree | 2 |
The warranty period is 1 year from the sale date, excluding cases of damage due to improper handling or non-compliance with manufacturer’s installation instructions.
The auxiliary power supply is 100-240V AC+10-15% (50-60Hz) or 100V DC+40-25%.
The torque should be about 1.47 - 1.96Nm.
The ambient temperature should not exceed 50°C.