Product leaflet for the Mettler Toledo M100 DR, a 2-wire HART analytical transmitter for pH/redox, pH/pNa, oxygen and conductivity measurements.
The Mettler Toledo M100 DR is a compact 2-wire DIN rail transmitter with HART communication for pH/redox, pH/pNa, oxygen and conductivity measurements. It supports ISM probes, Plug and Measure functionality, and probe diagnostics through HART.
This product leaflet provides dimension diagrams, technical characteristics, terminal definitions, LED functions, ordering references and a parameter selection guide for the M100 DR. This manual is written in French.
| HART communication | Standard HART communication |
|---|---|
| Mounting | 35 mm DIN rail systems |
| Width | 22.5 mm |
| Analog output | 4 to 20 mA with HART |
| Supply voltage | 14 to 30 V DC |
| Ambient operating temperature | -20 to 60 °C |
| Maximum probe cable length | 80 m |
| Enclosure classification | IP 20 |
| Weight | 0.5 kg |
| Galvanic isolation | Up to 500 V between inputs, outputs and ground |
It measures pH/redox, pH/pNa, oxygen and conductivity. It is described as a single-channel, multiparameter transmitter.
Yes. HART communication is standard, and configuration can be done through a HART handheld terminal or other HART asset management tools.
The supply voltage range is 14 to 30 V DC.
It is suitable for 35 mm DIN rail systems. The compact support is 22.5 mm wide.