KROHNE GS 8B calibrated flow simulator data sheet for electromagnetic flowmeters: 5 selectable ranges, zero flow simulation, specs and compatible converters.
The KROHNE GS 8B is a calibrated flow simulator that provides calibrated feedback to an electromagnetic flowmeter converter, simulating specific calibrated flow conditions in place of the meter flow tube. It offers 5 selectable simulated flow range settings plus a zero flow position, and can be used with many KROHNE 4-wire converters. The GS 8B is a passive device that derives its operating power from the converter and delivers a precision signal back to it; a zero position on the range switch allows determination of any zero offset present. Each instrument is factory calibrated to verify the accuracy of its operation.
This data sheet covers the system and operational description, the KROHNE converters the simulator can be used with (including IFC 300 C/F/W, IFC 010, IFC 020, IFC 090, IFC 110, IFC 080, SC 100 A/AS, SC 80 A/AS, T 900 F, 900 F/NF, T 900 E, F 200 and IFC 110 PF & SC 100 PF for full pipe simulation) and the older converters it is not suitable for (IFC 040, IFM 4042, TIV 50, TIV 60, SC 150). It also lists the included items, such as the connecting cables M, Z and Z3, carry/storage bag, technical data and operating instructions, calibration certificate and a CD, and gives the technical data with an annual recommended calibration interval.
| Ambient Temperature | -4°F to +140°F |
|---|---|
| Protection Category | IP 40, per IEC/EN 60529 |
| Field Current | ±60 to ±125 mA pulsed DC, 120 to 250 mA peak to peak |
| Error (% of measured value) | ±0.4%; SC 80 A: ±0.6% |
| Recommended Calibration Interval | Annual |
| Simulated flow range settings | 5 selectable settings plus zero flow |
It provides calibrated feedback to an electromagnetic flowmeter converter, simulating specific calibrated flow conditions in place of the meter flow tube.
It is not suitable for the older IFC 040, IFM 4042, TIV 50, TIV 60 and SC 150 signal converters.
No. Like the meter flow tube it is a passive device that derives its operating power from the converter.
The recommended calibration interval is annual.