Operation manual for the JENCO 6173 microcomputer-based pH/mV/temperature bench meter, covering setup, calibration, measurement, and AUTOLOCK operation.
The JENCO 6173 is a microcomputer-based bench meter for measuring pH, mV, and temperature. It provides AUTOLOCK and user-prompting features for pH and mV measurements, automatic or manual temperature compensation, and electrode efficiency display.
The manual covers initial inspection, battery replacement, LCD and key functions, connectors, buffer selection, pH calibration with ATC or manual temperature compensation, mV measurements, and pH buffer temperature coefficients.
| Measurement functions | pH, mV, and temperature |
|---|---|
| mV measurement range | -2000 to +2000 mV |
| Power supply | Six 1.5 V AA batteries |
| Maximum displayed temperature | Less than 60 °C |
| Available pH buffer sets | 4.01/7.00/10.01 pH and 4.00/6.86/9.18 pH |
| AUTOLOCK activation (mV) | Display changes less than 1 mV within 10 seconds |
It operates with six 1.5 V AA batteries.
The meter measures from -2000 mV to +2000 mV.
Turn off the meter, then press and hold the STAND key while turning it on again.
The meter can operate for about one hour after the BAT indicator appears.