JENCO 6173 pH/mV/Temperature Bench Meter Operation Manual
The JENCO 6173 is a microcomputer-based bench meter for measuring pH, mV, and temperature. It provides AUTOLOCK and user-prompting features for pH and mV measurements, automatic or manual temperature compensation, and electrode efficiency display.
The manual covers initial inspection, battery replacement, LCD and key functions, connectors, buffer selection, pH calibration with ATC or manual temperature compensation, mV measurements, and pH buffer temperature coefficients.
JENCO 6173 specifications
| Measurement functions | pH, mV, and temperature |
|---|---|
| mV measurement range | -2000 to +2000 mV |
| Power supply | Six 1.5 V AA batteries |
| Maximum displayed temperature | Less than 60 °C |
| Available pH buffer sets | 4.01/7.00/10.01 pH and 4.00/6.86/9.18 pH |
| AUTOLOCK activation (mV) | Display changes less than 1 mV within 10 seconds |
Source: JENCO 6173 pH/mV/Temperature Bench Meter Operation Manual
All JENCO 6173 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How many batteries does the JENCO 6173 use?
It operates with six 1.5 V AA batteries.
From: JENCO 6173 pH/mV/Temperature Bench Meter Operation Manual
What is the mV measurement range?
The meter measures from -2000 mV to +2000 mV.
From: JENCO 6173 pH/mV/Temperature Bench Meter Operation Manual
What buffer sets are available for pH calibration?
The meter supports two buffer sets: 4.01, 7.00, 10.01 pH and 4.00, 6.86, 9.18 pH.
From: Jenco Model 6173 Microcomputer Based pH/mV/Temperature Bench Meter Operation Manual
How does the AUTOLOCK feature work for pH measurements?
In AUTOLOCK mode, the instrument waits for a stable reading, then locks the display to the buffer value corresponding to the temperature.
From: Jenco Model 6173 Microcomputer Based pH/mV/Temperature Bench Meter Operation Manual
What calibration solutions are supported?
The device supports two sets of calibration solutions: 7.00, 4.01, 10.01 pH or 6.86, 4.00, 9.18 pH.
Does the 6173R model have a computer interface?
Yes, the 6173R model can connect to a computer via RS232C for monitoring and recording data.