JENCO Model 3331 Conductivity/Resistivity Transmitter User Manual

JENCO Model 3331 microcomputer conductivity/resistivity transmitter user manual covering installation, operation, calibration, relay control, errors, and…


JENCO Model 3331 Conductivity/Resistivity Transmitter User Manual - cover page
Brand
JENCO
Category
Other device
Document type
User Manual
Language
Chinese
Pages
26
File format
PDF
File size
226 KB
Published
10 November, 2026
Updated
10 November, 2026
Model
3331
MD5 checksum
1B6AC2DD1DB90285D5E6B1E6A32C02C2

About this manual

JENCO Model 3331 is a microcomputer-based conductivity/resistivity transmitter in a 1/8 DIN housing for laboratory and control applications. It can display conductivity, resistivity, or temperature, performs self-diagnosis at power-on, and provides one isolated current output with a user-settable range.

The manual covers installation, front-panel keys and display, terminal wiring, measurement, setting, and conductivity/resistivity calibration modes, relay control, error displays and causes, specifications, and warranty. This manual is written in Chinese.

What's inside

  • Installation steps
  • Front panel and keys
  • Display
  • Terminal wiring diagram
  • Measurement mode
  • Setting mode
  • Conductivity/resistivity calibration mode
  • Relay control
  • Error display and causes
  • Specifications

Specifications

Power supply100-240 VAC, 50/60 Hz
Ambient temperature0.0 to 50.0 °C
EnclosureIP65, 1/8 DIN, depth 90 mm
Weight290 g
Temperature range-10.0 to 120.0 °C
Temperature resolution0.1 °C
Conductivity/resistivity accuracy±0.5% full scale ±1 digit
Reference temperature25.0 °C
Temperature coefficient2.00% or 0.00%, user selectable
Electrode constants0.01, 0.1, 1.0 (2-wire conductivity electrode)
Temperature sensorThermistor, 10 kΩ at 25 °C
4-20 mA output range4 to 20 mA (isolated)
Maximum load500 Ω
Current accuracy±0.03 mA
Isolation voltage500 VDC

Frequently asked questions

How do I enter setting mode on the Model 3331?

In measurement mode, press DOWN and ENTER together.

How do I enter calibration mode?

In measurement mode, press UP and ENTER together.

What power supply does the transmitter require?

100 to 240 VAC at 50/60 Hz.

What is the maximum load for the 4-20 mA output?

500 ohms.

Download PDF (226 KB)