Getinge urgent field safety notice for Datascope IABP models covers battery usage, charging, maintenance, storage, and interim user actions.
Getinge issued an urgent field safety notice for Datascope Intra-Aortic Balloon Pump (IABP) models, including Cardiosave Hybrid, Cardiosave Rescue, CS300, and CS100. The notice reminds users to follow each device's operating instructions for battery usage, charging, maintenance, and storage because battery run times and discharge cycles vary by model.
Improper battery maintenance may reduce expected battery run time. The notice lists interim actions such as keeping the IABP plugged into AC power, charging batteries before transport, checking run time, and replacing batteries when required. It also describes corrective actions, including battery operation guides and planned software upgrades.
| Cardiosave battery labeled lifetime | 4 years |
|---|---|
| CS100/CS300 battery discharge cycle threshold | 100 total discharge cycles |
| Maximum balloon inactivity time | 30 minutes |
The affected models are Cardiosave Hybrid IABP, Cardiosave Rescue IABP, CS300 IABP, and CS100 IABP.
Transfer the patient to an alternative Datascope IABP; if one is unavailable, manually inflate the IAB with air or helium and immediately aspirate, or remove the IAB.
They should be replaced immediately because their labeled lifetime is 4 years.
It indicates that the IABP internal battery requires maintenance.