This document describes how to use the Cardiohelp Blood Perfusion System, including preparation and installation, cleaning the system, connecting to the patient, starting and ending perfusion, etc., mainly for blood perfusion
The Cardiohelp System Setup Guide details the procedures for preparing and installing the Cardiohelp blood perfusion system. It covers the connection of integrated sensors, flow/bubble sensors, and gas tubes, as well as the installation of the HLS Module and Prime Bag. The guide also explains how to connect the system to the patient using sterile techniques and how to manage the tubing connections between the Prime Set and the Table Set.
| Prime Volume | 1.5 to 2 l |
|---|---|
| Minimum Height | 60 cm (2 ft) |
| Priming Speed | 3000 rpm |
| De-airing Speed | 4000 rpm |
You must add 1.5 to 2 liters of Physiological Solution via the spike of the Prime Line.
There must be at least 60 cm (2 ft) between the top of the Cardiohelp and the bottom of the Prime Bag.
Set the speed to 3000 rpm for 2 minutes to passively prime the set.
Set the speed to 4000 rpm for 1 minute to de-air the 3 Luer lock connectors.