This is an operation guide that introduces how to use Cardiosave IABP for transport operation.
This document serves as an abbreviated quick reference guide for the Getinge Cardiosave Intra-aortic Balloon Pump (IABP) system, specifically for software version C.06 intended for use outside the United States. It provides essential instructions for initial setup using both fiber-optic and conventional IAB catheters, including verifying alarms and automatic calibration features.
The guide details procedures for portable operation, including battery and helium reservoir checks, as well as the physical steps for removing the pump console from the hospital cart for transport and re-inserting it upon return. It also covers sizing guidance for the Intra-aortic Balloon Sizing Guide based on patient height.
A fully charged internal helium reservoir will provide approximately 36 fill cycles.
An operator may invoke a calibration anytime by pressing and holding the CALIBRATE PRESSURE key for 2 seconds.
Three audio tones will sound when the console is locking into the cart.