This manual provides instructions for setting up, operating, and verifying the performance of the Fluke 5320A-LOAD High Voltage Load Adapter.
The document is a user guide for the Fluke 5320A-LOAD High Voltage Load Adapter. It describes how to connect the adapter to the Fluke 5320A Electrical Safety Analyzer and explains safety precautions regarding high voltage, grounding, and ventilation.
The manual details procedures for using the adapter to test leakage currents, including selecting appropriate resistors based on voltage levels. It also outlines two methods for verifying the electrical performance of the load: a voltage method using a precision power supply and a multimeter measurement method. This manual is written in Chinese.
| Resistance Range | 10 kΩ to 5 MΩ |
|---|---|
| Operating Temperature | 5 °C to 40 °C |
| Storage Temperature | -20 °C to +70 °C |
| Tolerance | 10% |
Under normal operating conditions, the resistance of the load should be checked at least once a year.
The fuse should be replaced only with a 250V fuse of the indicated rating.