Operating instructions for the Elcometer 236 DC Holiday Detector, covering safe high-voltage coating flaw inspection, setup, use, troubleshooting, and…
The Elcometer 236 DC Holiday Detector locates flaws in non-conducting coatings on conducting substrates, including coatings up to 7 mm thick. It applies high DC voltage through a probe and connects to the substrate via a high voltage return lead; a spark, audible alarm, and visual alarm indicate a flaw.
The manual covers safe use, standards and test methods, setup, operation, static electricity, current monitoring, voltage and sensitivity settings, probe selection, troubleshooting, maintenance, storage, spares, technical specification, and related equipment. Safety guidance includes not using the detector in combustible atmospheres, in rain or damp conditions, or while the charger is connected; the battery is supplied discharged and must be charged before first use.
| High voltage at probe | Up to 30,000 volts |
|---|---|
| Maximum coating thickness | Up to 7 mm (275 mils) |
| Minimum coating thickness | At least 200 µm |
| Preferred coating thickness | Over 500 µm |
| Probe speed (ASTM D 4787) | 0.3 m/s (1 ft/s) maximum |
| RF emissions at 3 m | Approximately 60 dBµV/m from 30 MHz to 1000 MHz |
| Distance from sensitive electronic equipment | Not within 30 m |
It can test coatings up to 7 mm (275 mils) thick and requires coatings at least 200 µm thick, preferably over 500 µm thick.
No, it must not be used for high voltage testing while connected to the mains electricity supply for battery charging.
Wear rubber gloves, work well clear of other personnel, and do not use the equipment in combustible or flammable atmospheres.
For safety and reliability, the detector is supplied with the battery discharged and it must be recharged before first use.