This manual provides operating instructions for the Elcometer 5100 Payne Permeability Cup. The manual describes how to operate the instrument and how to perform measurements.
The Elcometer 5100 Payne Permeability Cup is a simple instrument used to determine the vapour transmission rate of unsupported films of paint, varnish, plastic, paper, and other materials. Manufactured from anodised aluminium, it is available in three sizes to suit different film testing requirements. The device operates by clamping a film over the cup, weighing the assembly before and after a set period, and calculating the water-vapour transmission rate based on the mass difference.
| Material | anodised aluminium |
|---|---|
| Model 5100/1 Volume | 15 cm³ |
| Model 5100/2 Volume | 50 cm³ |
| Model 5100/3 Volume | 75 cm³ |
| Model 5100/1 Area | 10 cm² |
| Model 5100/2 Area | 30 cm² |
| Model 5100/3 Area | 30 cm² |
The instrument can be used in accordance with ASTM D1653, ASTM E96, and ISO 7783.
The rate is calculated using the formula P = 240 x Dm / A, where Dm is the rate of change in mass in mg/h and A is the area in square centimetres.
The box contains the Elcometer 5100 Payne Permeability Cup, a storage case, and operating instructions.