A command and procedure guide for Bruker TopSpin NMR software, covering basic commands, calibration, data processing, and 1D, 2D and 3D experiment setup.
This Bruker TopSpin NMR software guide lists basic TopSpin commands for NMR work, with comments on sample handling, locking, temperature setting, tuning and matching, shimming, experiment setup, acquisition, and data processing.
It also gives common procedures for temperature calibration, 1D 1H measurement setup, selective pulse calibration for solvent flip-back, and 2D and 3D experiment setup and measurement.
The edc command is listed for creating a new experiment.
The zg command is used to go and measure data.
calctemp is used after processing the methanol spectrum in the temperature calibration procedure.
The procedure lists efp for 1D processing, followed by phase correction if necessary using .ph and .sret.