The Bruker AQR ASU2 / Local Oscillator & Tune User Manual provides wiring diagrams, interface connections, and full specifications for the instrument.
This manual describes the Bruker AQR ASU2 / Local Oscillator & Tune, including front‑panel wiring, functional block diagrams, RF wiring, and interface connector pinouts. It also lists the electrical and performance specifications required for installation and operation.
Key sections cover the device description, detailed wiring diagrams, connector tables, and a comprehensive specifications table covering frequency range, gain, power levels, VSWR, dynamic range, isolation, and noise performance.
| Frequency range | 5 to 500 MHz |
|---|---|
| Gain | 1 dB ± 1 dB |
| Input power | 4 dBm ± 0.5 dB |
| Output power | 5 dBm ± 1 dB (for 4 dBm IN) |
| Input VSWR | 1.4 |
| Output VSWR | 1.5 |
| Dynamic range | 110 dB |
| Modulation dynamic range | 50 dB |
| Multiplication dynamic range | 30 dB |
| Isolation (input/output) | > 70 dB (up to > 110 dB in some configurations) |
| Output noise level (unblanked, MOD/MULT = 0 V) | < ‑135 dBm (1 Hz) |
The instrument operates from 5 MHz up to 500 MHz.
The specified dynamic range is 110 dB.
Isolation between input and output is greater than 70 dB, with some configurations exceeding 110 dB.
The unblanked output noise level is below –135 dBm (1 Hz).