This Bruker AssureSST user manual explains how to configure and run system suitability tests for NMR spectrometer performance using IconNMR and TopSpin.
AssureSST runs a user-selected set of experiments to monitor instrument performance. It is configured and controlled through the IconNMR automation software package, which uses TopSpin to acquire and evaluate data, and generates a PDF report summarizing the specifications tested and results obtained.
Configuration is done in the IconNMR Configuration window using the command “iconc” from the TopSpin command line. The AssureSST section has three tabs: System Suitability Test (SST), SST Standard Tests, and SST User Tests. The SST includes up to six standard experiments and up to four user-defined experiments, and must pass before queued samples are acquired.
| Standard experiments | up to six |
|---|---|
| User-defined experiments | up to four |
AssureSST parameters are accessed through the IconNMR Configuration window using the command “iconc” from the TopSpin command line.
A PDF report is generated summarizing the specifications tested and the results obtained.
Yes, the SST as a whole must pass before any other samples queued in IconNMR can be acquired.