Bruker AssureSST System Suitability Test User Manual

This Bruker AssureSST user manual explains how to configure and run system suitability tests for NMR spectrometer performance using IconNMR and TopSpin.


Bruker AssureSST System Suitability Test User Manual - cover page
Brand
Bruker
Category
Other Electronics
Document type
User Manual
Language
English
Pages
18
File format
PDF
File size
2.4 MB
Published
09 November, 2026
Updated
09 November, 2026
MD5 checksum
18FC39DF44E7E2DCB1AE1F2DD11F3DDC

About this manual

AssureSST runs a user-selected set of experiments to monitor instrument performance. It is configured and controlled through the IconNMR automation software package, which uses TopSpin to acquire and evaluate data, and generates a PDF report summarizing the specifications tested and results obtained.

Configuration is done in the IconNMR Configuration window using the command “iconc” from the TopSpin command line. The AssureSST section has three tabs: System Suitability Test (SST), SST Standard Tests, and SST User Tests. The SST includes up to six standard experiments and up to four user-defined experiments, and must pass before queued samples are acquired.

What's inside

  • Overview
  • AssureSST Configuration
  • System Suitability Test (SST) Tab
  • SST Standard Tests Tab
  • ¹H Lineshape
  • ¹H Sensitivity
  • ¹³C Sensitivity

Specifications

Standard experimentsup to six
User-defined experimentsup to four

Frequently asked questions

How is AssureSST configured and controlled?

AssureSST parameters are accessed through the IconNMR Configuration window using the command “iconc” from the TopSpin command line.

What report is generated after the SST?

A PDF report is generated summarizing the specifications tested and the results obtained.

Must the SST pass before queued samples are acquired?

Yes, the SST as a whole must pass before any other samples queued in IconNMR can be acquired.

Download PDF (2.4 MB)

More Bruker Manuals


View all Bruker manuals →