Berthold LB 6500-H10 GM dose rate probe datasheet covering technical data for radiation protection dose rate measurement.
This Berthold datasheet describes the LB 6500-H10 GM dose rate probe for ambient equivalent dose rate measurement in radiation protection. It is used for environmental monitoring or local dose rate measurements and is available as the LB 6500-4-H10 low-dose probe (≤ 10 mSv/h) and the LB 6500-3-H10 high-dose probe (≤ 1 Sv/h).
The probe contains an energy-compensated GM tube with amplifier, integral discriminator, and high-voltage generator, outputting digital pulses proportional to dose rate. A 7-core shielded cable 10 m long is permanently attached (extendable to ≤ 100 m), with an optional Tuchel connector variant. Operating voltage is 450 V or 575 V depending on version, with IP65 protection and dimensions of 45 mm diameter by 315 mm length. This manual is written in German.
| Dose rate range LB 6500-4-H10 | 100 nSv/h – 10 mSv/h |
|---|---|
| Energy range LB 6500-4-H10 | 65 keV – 1,3 MeV ± 40% |
| Operating voltage LB 6500-4-H10 | 450 V |
| Current consumption | 15 mA |
| Ambient temperature | -20 to +40 °C |
| Cable length | 10 m (max. 100 m) |
| Weight | ca. 1000 g |
It is used for dose rate measurement in radiation protection, for environmental monitoring or local dose rate measurements.
LB 6500-4-H10 is a low-dose probe up to 10 mSv/h, and LB 6500-3-H10 is a high-dose probe up to 1 Sv/h.
The operating voltage is 450 V for LB 6500-4-H10 and 575 V for LB 6500-3-H10.