Leaflet for the Berthold CentroLIA pc LB 962 microplate luminometer, covering assays, detector, plate formats, software and order information.
The CentroLIA pc LB 962 is a microplate luminometer from BERTHOLD TECHNOLOGIES intended for luminescence immunoassays as well as reporter gene, ATP, caspase and kinase assays. It uses a low-noise photomultiplier tube in single-photon counting mode, reaches a stated sensitivity of 0.03 mIU/L TSH and a dynamic range of more than six orders of magnitude, and reads 96 well opaque microplates and 8 well or 12 well strip plates through a USB interface to a PC.
The accompanying software provides protocols for immunometric (ILMA), competitive (LIA), cut-off and raw data assays, with standard curve calculation by cubic spline, linear regression or 4PL and log/log or logit/log axis transformation, plus master and reference curve handling. The document also lists operation modes, PC requirements, power supply, temperature and humidity ranges, dimensions, weight and the order information for the instrument and a luminescence test plate.
| Detector | Low-noise photomultiplier tube in single-photon counting mode |
|---|---|
| Sensitivity | 0.03 mIU/L TSH; 20 amol ATP; ~7.5 zmol firefly luciferase |
| Dynamic range | > 6 orders of magnitude |
| Crosstalk | < 10-6 |
| Plate formats | 96 well opaque microplates and strip plates, 85.48 x 127.76 x 14 mm |
| Interface | USB |
| PC operating system | Win 2000, Win NT, Win XP, Win Vista |
| Power supply | 110-240 V, 50/60 Hz, 30 VA |
| Measurement time | 0.1 to 600 sec |
| Temperature range | Storage 0 - 40 °C; operation 15 - 35 °C |
| Dimensions | 300 x 400 x 200 mm (W x D x H) |
| Weight | 6 kg |
| Regulations | CE, UL, CSA |
| Order number | CentroLIA pc LB 962 52180-10 |
It measures 96 well opaque microplates and 8 well or 12 well strip plates with outer dimensions of 85.48 x 127.76 x 14 mm.
The stated dynamic range is more than six orders of magnitude.
Measurement time can be set from 0.1 to 600 seconds, with a delay of up to 600 seconds.
The document lists Win 2000, Win NT, Win XP and Win Vista.