Berthold BAI 9109-4 Beta-Gasmonitor product information: monitors radioactive gases in room and discharged air with 1 to 4 xenon-sealed detectors.
This product information sheet from Berthold Technologies describes the BAI 9109-4 Beta-Gasmonitor, a device for monitoring radioactive gases such as Krypton-85, Argon-41, Xenon-133, Carbon-11 and Fluorine-18 in rooms and discharged air, and for measuring beta- and beta+ (positron) nuclides. The measuring chamber can be equipped with 1 to 4 large area proportional counter tubes, xenon sealed, where one detector can be used as a compensation detector. Sample air is directed via a prefilter by means of a pump, and the chamber has a volume of 11.45 l with a lead shield on all sides of 2 cm thickness in 4π geometry.
The leaflet lists the minimal detectable activities in Bq/m³ according to DIN 25482-1 (measuring time 3600 s, background 6 cps) for 1 to 4 detectors, and names the data acquisition systems available depending on the application: Data Logger LB 112, Data Logger LB 5340 and Data Logger LB 9000. Technical data covers the measuring chamber, lead shield, detector, electronics and ambient conditions. The document is identified as BAI 9109-4 engl · Id. 48821PR2 · 08-2012 · Rev.0.
| Measuring chamber construction material | PVC |
|---|---|
| Measuring chamber volume | 11.45 l |
| Pump nominal flow rate through the vessel | max. 15 m³/h |
| Lead shield material | 96 % Pb + 4 % Sb |
| Lead shield geometry | 4π, all sided with 2 cm thickness |
| Lead shield weight | approx. 185 kg (lead only) |
| Detector type | sealed xenon large area proportional counter tubes LB 6350-1 |
| Sensitive detector area | 230 cm² |
| Background | typical approx. 4 cps per detector in 2 cm lead shield |
| Measuring range | 1 kBq/m³ to 100 MBq/m³ |
| Gamma sensitivity (Cs-137, 1 detector) | 110 cps per µSv/h |
| Electronics | Integrated preamplifier and high voltage supply LB 2022-22 |
| Ambient temperature | 0°C to +40°C |
| Sample air maximum differential pressure | 0.2 bar |
It monitors gases such as Krypton-85, Argon-41, Xenon-133, Carbon-11 and Fluorine-18 in rooms and discharged air.
The chamber can be equipped with 1 to 4 large area proportional counter tubes, xenon sealed, and one detector can serve as a compensation detector.
The measuring chamber has a volume of 11.45 l.
Data Logger LB 112, Data Logger LB 5340 and Data Logger LB 9000 are available depending on the application.