Instructions for use of the Beckman Coulter S2096 swinging-bucket rotor for Allegra X-22, Allegra 21, GS-15, Spinchron 15 and Allegra X-30 centrifuges.
This instruction manual (GS-TB-005GA, April 2011) covers the Beckman Coulter S2096 two-place swinging-bucket rotor, for use in the Beckman Coulter Allegra X-22 Series, Allegra 21 Series, GS-15 Series, Spinchron 15 Series, and Allegra X-30 Series centrifuges. It contains a safety notice, specifications, a description of the rotor and carriers, preparation and use, prerun safety checks, installing the rotor yoke, loading the carrier, operation, removal and sample recovery, labware, run speeds, care and maintenance, and storage.
The rotor yoke is made of aluminum and is black-anodized for corrosion protection, and the black-anodized aluminum carriers swing out to horizontal position during centrifugation. Each carrier can hold standard microplates for serial dilution of small liquid volumes — up to three stacked multiwell plates or one deepwell or square-well plate. The manual closes with returning a rotor, the supply list and replacement rotor parts, and the Beckman Coulter, Inc. benchtop rotor warranty.
| Maximum speed | 3000 rpm |
|---|---|
| Critical speed range | 500 to 1300 rpm |
| Density rating at maximum speed | 1.2 g/mL |
| Maximum load per bucket | 320 grams |
| Maximum allowable imbalance of opposing loads | 10 grams |
| Weight of fully loaded rotor | 2.06 kg (4.54 lb) |
| Rotor and lid material | Aluminum |
| Approximate acceleration time to maximum speed (fully loaded) | 21 sec |
| Rotor type | Two-place swinging-bucket rotor |
| Carrier capacity | Up to three stacked multiwell plates, not to exceed 40.6 mm (1.6 in.) deep, or one deepwell or square-well plate |
It is intended for the Beckman Coulter Allegra X-22 Series, Allegra 21 Series, GS-15 Series, Spinchron 15 Series, and Allegra X-30 Series centrifuges.
The maximum speed is 3000 rpm, with a density rating at maximum speed of 1.2 g/mL.
Hook both carriers to the rotor for every run, and make sure filled containers are loaded symmetrically with opposing labware filled to the same level.
Beckman Coulter agrees to correct defects of material or workmanship that develop within seven years after delivery of a benchtop centrifuge rotor to the original buyer.