Beckman Coulter PA 800 plus Pharmaceutical Analysis System Pre-Installation Guide

Pre-installation guide for the Beckman Coulter PA 800 plus Pharmaceutical Analysis System with 32 Karat Software Version 9.1 and PA 800 plus Software…


Beckman Coulter PA 800 plus Pharmaceutical Analysis System Pre-Installation Guide - cover page
Brand
Beckman Coulter
Category
Analyzers
Document type
Installation Guide
Language
English
Pages
28
File format
PDF
File size
394 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
969AF6AB672CDCEF18E06DDC567E7DA9

About this manual

This guide provides pre-installation information for the Beckman Coulter PA 800 plus Pharmaceutical Analysis System, including software versions 32 Karat 9.1 and PA 800 plus 1.1. It covers safety notices, system specifications, computer requirements, detector specifications, and training requirements.

The specifications include system weight, power, heat dissipation, environmental limits, sound pressure, and requirements for UV, PDA, and LIF detectors. Training information states that up to two operators can be trained after installation.

What's inside

  • Safety Notices
  • Symbols and Labels
  • Revision History
  • Specifications
  • PA 800 plus System
  • Computer Requirements
  • UV Detector
  • 3 PDA Detector
  • LIF Detector
  • PA 800 plus System Training

Specifications

Weight (PA 800 plus)188 lbs (85.3 kg)
Power requirements (PA 800 plus)100-240VAC, 5.0A, 50/60Hz; auto-voltage sensing power supply
Maximum heat dissipation (PA 800 plus)400 W (1024 BTUs/Hour)
Ambient operating temperature range15-40°C (15-30°C recommended)
Humidity restrictionsRH<80% (non-condensing) @ 30°C
Altitude restrictionsup to 2000m (6,562 ft)
Maximum sound pressure70 dB
Maximum sound pressure at 1 meter away66 dB
Computer CPUIBM PC (currently validated on IBM M52 and M57 models)
Computer memory1024 Mb (minimum)
Computer hard drive40 Gb or larger
Operating systemMicrosoft Windows XP Pro with Service Pack 2
UV detector wavelength range190 to 600 nm
UV detector sourceDeuterium Lamp; 30 W; prealigned
PDA detector256 Element Diode Array
PDA detector bandwidth6 nm minimum (Absorbance Averaging)
LIF detector RFU range0 to 1000 RFU
LIF detector sensitivity1 x 10-11M Sodium Fluorescein with a signal-to noise ratio > 2
External laser module fiber cable length6 feet (1.83 m)
LIF software requirements32 Karat Software, Version 9.0 or later

Frequently asked questions

What software versions does this guide cover?

It covers 32 Karat Software Version 9.1 and PA 800 plus Software Version 1.1.

How many operators can be trained at one time?

No more than two operators can be trained on the instrument at any given time.

What operating system is required for the PC?

Microsoft Windows XP Pro with Service Pack 2 is required.

What is the weight of the PA 800 plus?

The PA 800 plus weighs 188 lbs (85.3 kg).