User manual for the Bante Instruments PHscan 20S pen-style pH meter covering operation, calibration, measurement, maintenance, and troubleshooting.
The PHscan 20S is a portable pen-style pH meter designed for laboratory or field pH measurement. It features a clear LCD display with mode icons, automatic temperature compensation (ATC), and a simple membrane keypad for power, calibration, setup, and data hold functions.
The manual details startup, electrode preparation, buffer solution preparation, setup menu configuration (buffer standard, calibration points, auto-hold, auto-power-off, factory reset), 1- to 3-point calibration procedures using USA or NIST buffers, measurement and data hold modes, electrode cleaning and maintenance guidelines, sensor and battery replacement steps, troubleshooting error codes, and full technical specifications. This manual is written in Chinese.
| Model | PHscan20S |
|---|---|
| pH measurement range | -1.00~15.00 pH |
| pH measurement accuracy | ±0.01 pH |
| Calibration points | 1 to 3 points |
| Calibration solutions | USA (pH4.01/7.00/10.01) or NIST (pH4.01/6.86/9.18) |
| Temperature compensation range | 0~60 °C (32~140 °F) |
| Temperature compensation method | Automatic |
| Data hold | Manual or automatic endpoint hold |
| Power off mode | Manual or automatic (after 8 minutes no key operation) |
| One-key reset | Included |
| Sensor type | E-PHSCAN-ST |
| Operating temperature | 0~60 °C (32~140 °F) |
| Power supply | 2 × 1.5V AAA batteries |
| Dimensions | 185 (L) × 40 (diameter) mm |
| Weight | 100 g |
It supports USA standard buffers (pH 4.01, 7.00, 10.01) and NIST standard buffers (pH 4.01, 6.86, 9.18).
The electrode sensor is too dry; immerse it in tap water or 3M KCl solution for 10 minutes to rehydrate.