This document is a user manual for the DOscan10 pen-type dissolved oxygen meter from Shanghai Bante Instruments Co., Ltd. The document covers topics such as battery installation, key functions, display screen icons, electrolyte filling, electrode polarization, and setup menu.
The Bante Instruments DOscan10 is a portable pen-type dissolved oxygen meter designed for measuring dissolved oxygen in water, wastewater, saltwater, and other liquids. The device features a custom LCD display, automatic temperature compensation (ATC), and the ability to compensate for atmospheric pressure and salinity. This manual is written in Chinese.
| Measurement Range (DO) | 0.0 to 20.0 mg/L |
|---|---|
| Display Resolution (DO) | 0.1 mg/L |
| Measurement Accuracy (DO) | ±0.5 mg/L |
| Salinity Compensation Range | 0 to 35 g/L |
| Operating Temperature | 0 to 50° C, 32 to 122° F |
| Waterproof Rating | IP54 |
| Power Supply | 2 x 1.5V AAA batteries |
The electrode must be polarized for 10 minutes by turning on the meter and holding the Meas key.
A two-point calibration requires using air-saturated water for the first point and zero oxygen solution for the second point.