The BI-620 industrial pH controller is a powerful instrument. This user manual provides a step-by-step guide to help users operate the meter correctly. Before installation, ensure that the environmental conditions meet the requirements, and pay attention to safety precautions during the installation process.
The Bante Instruments BI-620 is an industrial pH controller designed for measuring pH and mV values. The device features a setup menu for customizing display options, alarm limits, and communication protocols, as well as automatic temperature compensation (ATC). It supports single-point, two-point, or three-point calibration using standard USA or NIST buffer solutions.
| Relative Humidity | less than 80% |
|---|---|
| Ambient Temperature | 5°C (41°F) and 50° C (122° F) |
| Power Supply | DC 24V |
| Analog Output | 4 to 20 mA |
| Communication Protocol | RS485 |
| Calibration Points | up to 3 points |
Installation requires relative humidity less than 80%, ambient temperature between 5°C and 50°C, no electromagnetic interference, and no corrosive gas.
The meter allows up to three points of pH calibration, though at least two points are recommended for high accuracy.
It recognizes USA standard buffers (pH 4.01, 7.00, 10.01) and NIST standard buffers (pH 4.01, 6.86, 9.18).
If the calculated electrode slope falls outside this range during calibration, the pH electrode may need to be replaced.