Agilent 1290 Infinity II LC Specification Compendium

A specification compendium detailing physical and performance specifications for Agilent 1290 Infinity II Liquid Chromatography components including pumps…


Agilent 1290 Infinity II LC Specification Compendium - cover page
Brand
Agilent Technologies
Document type
Datasheet
Language
English
Pages
38
File format
PDF
File size
691 KB
Published
16 October, 2026
Updated
16 October, 2026
MD5 checksum
BB0BFAA61CE15E6CFDFFB8631A766C23

About this manual

This document serves as a technical specification guide for the Agilent 1290 Infinity II LC system family. It provides detailed data on physical dimensions, environmental requirements, and safety standards for key modules such as the Flexible Pump (G7104A) and High Speed Pump (G7120A).

The compendium covers performance metrics including flow precision, pressure ranges, composition accuracy, and gradient formation capabilities. It also includes specifications for injectors like the Multisampler (G7167B), UV detectors, special detectors such as the Evaporative Light Scattering Detector (ELSD), and the Multicolumn Thermostat.

What's inside

  • Agilent 1290 Infinity II Flexible Pump
  • Agilent 1290 Infinity II High Speed Pump
  • Agilent 1290 Infinity II Multisampler
  • Agilent 1290 Infinity II Variable Wavelength Detector
  • Agilent 1290 Infinity II DAD FS
  • Agilent 1290 Infinity II Evaporative Light Scattering Detector
  • Agilent 1290 Infinity II Multicolumn Thermostat

Specifications

Weight (Flexible Pump)16.1 kg (35.5 lbs)
Dimensions (Flexible Pump)180 x 396 x 436 mm
Flow Range (Flexible Pump)0.001 – 5 mL/min
Pressure Range (Flexible Pump)up to 130 MPa at 0 – 2 mL/min
Weight (High Speed Pump)21.0 kg (46.3 lbs)
Flow Precision (High Speed Pump)≤0.07 % RSD or 0.005 min SD
Weight (Multisampler)22 kg (48.5 lbs) w/o sample cooler

Frequently asked questions

What is the maximum operating altitude for the Agilent 1290 Infinity II Flexible Pump?

The pump can operate at altitudes up to 3000 m (9842 ft).

What is the recommended pH range for the Agilent 1290 Infinity II Flexible Pump?

The recommended pH range is 1.0 to 12.5.

What is the settable flow range for the Agilent 1290 Infinity II High Speed Pump?

The settable flow range is 0.001 to 5 mL/min.

Download PDF (691 KB)