testo 110 temperature measuring instrument user manual, introduces product safety instructions, intended use, product description, debugging, operation, measurement, maintenance and maintenance, questions and answers, technical data, accessories/spare parts and other content
This manual provides safety guidelines, operational steps, and maintenance instructions for the Testo 110 temperature measuring instrument. It covers connecting probes, setting alarms, and using the infrared and radio interfaces. The device is designed for food industry and laboratory use, and must not be used in explosive areas or for medical diagnostics. Technical specifications include measurement ranges, accuracy, and power requirements. This manual is written in Chinese.
| Measuring Range (NTC Probe) | -50.0...+150.0°C / -58.0...+302.0°F |
|---|---|
| Accuracy (NTC Probe) | ±0.2°C / ±0.4°F (-20.0...+80.0°C / -4.0...+176.0°F) ±0.3°C / ±0.6°F (rest of range) |
| Operating Temperature Range | -20 ... +50°C / -4 ... +122°F |
| Voltage Supply | 1x 9 V monobloc battery/rech. battery |
| Running Time (with Probe Connected) | approx. 200h / approx. 68h (lighting off / on) |
Sensor is not plugged in. Switch instrument off, connect probe and switch instrument back on again.
The permitted measuring range for the NTC probe is -50.0...+150.0°C / -58.0...+302.0°F.
The instrument runs for approximately 200 hours with the display on and a probe connected.