This document describes the ATEX explosion-proof load cell, which complies with the ATEX Directive 94/9/EC.
This instruction manual from Zemic covers load cells certified for use in potentially explosive atmospheres under ATEX Directive 94/9/EC. It explains product labelling and identification, special conditions for safe use, installation requirements, and care of the load cells.
The document also describes maintenance tools and procedures, repair and warranty terms, and failure analysis with guidance on avoiding misuse. Load cell-specific information is stated to be printed on the calibration certificate.
| Certification | ATEX Directive 94/9/EC |
|---|---|
| Intended environment | Potentially explosive atmospheres |
| Function | Converts mechanical force into an electrical signal |
| Input/output impedance, 700 Ω bridge resistance | 1100–1410 Ω |
| Input/output impedance, 350 Ω bridge resistance | 500–710 Ω |
The load cells are certified according to ATEX Directive 94/9/EC.
Use shunt-diode barriers for load cell installation.
No. Fault location and troubleshooting should not be attempted in a hazardous area where there is danger of explosion.
Never use a Megohmmeter to measure input or output resistance.