This document provides cleaning, disinfection and sterilization instructions for Welch Allyn brand rigid reusable and single use sigmoidoscopes, anoscopes and accessories.
These instructions provide detailed procedures for the reprocessing of Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes, Anoscopes, and Accessories. The devices are used to illuminate and facilitate the examination of the rectal and anal cavity and may provide entry for surgical instruments.
The document categorizes endoscopes and the insufflation system as semi-critical devices requiring high-level disinfection, while the illumination system is categorized as a non-critical device requiring low-level disinfection.
| Autoclave Sterilization Temperature | 270° F (132° C) |
|---|---|
| High Level Disinfection Soak Time | Minimum 12 minutes |
The bulb must be dried sufficiently, which can be achieved by hanging the bulb with the tube down for at least 24 hours.
No, the insufflation system must not be autoclaved as it will cause damage.
An enzymatic detergent formulated for endoscopic instruments is recommended; instructions were validated using ENZOL or CIDEZYME.