Welch Allyn DuraShock Non-Automated Sphygmomanometer Directions for Use

Welch Allyn non-automated sphygmomanometer directions for use covering intended use, operation, specifications, calibration, cleaning and warranty.


Welch Allyn DuraShock Non-Automated Sphygmomanometer Directions for Use - cover page
Brand
welchallyn
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
3.4 MB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
1604C048D620B9ACE91D9D7612A01254

About this manual

These directions for use apply to Welch Allyn DuraShock non-automated (aneroid) sphygmomanometer models including DS44, DS45, DS48, DS48A, DS54, DS55, DS56, DS58, DS65, DS66 and 767. The devices are noninvasive blood pressure measurement instruments intended for use with an inflatable cuff and stethoscope to determine blood pressure in pediatric through adult patients, and are contraindicated for neonate use.

The document covers cuff selection and placement, the inflation and deflation procedure, Korotkoff sound guidelines (K5 for adults, K4 for pediatric patients ages 3 to 12), technical and environmental specifications, calibration checks, cleaning and warranty terms.

What's inside

  • Intended use
  • Contraindications
  • Warnings and cautions
  • Operation
  • Operational Guidelines
  • Technical specifications
  • Environmental specifications
  • Calibration
  • Cleaning
  • Warranty

Specifications

Accuracy± 3 mm Hg
Operational Range0 to 300 mmHg
Displayed Range0 to 300 mmHg
Scale graduationIncrements of 2 mmHg
Pointer movementNo stop pin
Tube adapterIntegrated, 1, or 2 tube adapter
Pressure generationBulb
Pressure reductionAir release valve
Storage temperature-20 °C to 70 °C (-4 °F to 158 °F)
Operating temperature10 °C to 40 °C (50 °F to 104 °F)
Storage relative humidity15% to 95% (noncondensing)
Operating relative humidity15% to 90% (noncondensing)
Deflation rate2 to 3 mm Hg per second
Cuff position2.5 cm above the bend in the elbow
Inflation target30 mm Hg above estimated systolic pressure

Frequently asked questions

What is the intended use of Welch Allyn non-automated sphygmomanometers?

They are noninvasive blood pressure measurement devices intended for use with an inflatable cuff and stethoscope to determine blood pressure in pediatric through adult patients.

Can these sphygmomanometers be used on neonates?

No. The devices are contraindicated for neonate use, and safety and effectiveness with neonate cuffs (neo 1 to neo 5) is not established.

How often should the calibration be checked?

Welch Allyn recommends a full check of calibration at least every two years, or according to local law and after maintenance and repair.

Which accessories should be used with the device?

Use only Welch Allyn blood pressure cuffs and accessories; substitution may result in measurement error.