Sysmex XT 1800i Automated vs Manual Differential Comparison for Immature Granulocyte Alerts

This study compared the differences between SYSMEX XT1800i automatic classification and manual classification in immature granulocyte alerts. The results showed that the SYSMEX XT1800i automatic classification had a good correlation with the manual classification, but there were still some differences.


Sysmex XT 1800i Automated vs Manual Differential Comparison for Immature Granulocyte Alerts - cover page
Brand
Sysmex
Category
Professional medical equipment
Document type
Other
Language
Spanish
Pages
4
File format
PDF
File size
43 KB
Published
17 November, 2023
Updated
17 November, 2023
MD5 checksum
DC0ABD83DB0D4FB806CA2559612082B5

About this manual

This Spanish-language journal article compares the automated leukocyte differential on the Sysmex XT 1800i with a manual differential for immature granulocyte alerts. It reports 120 samples collected over eight weeks, a correlation coefficient of r = 0.55 between absolute values, and concludes that manual counting remains necessary for confirmation.

The article defines immature granulocytes as promyelocytes, myelocytes, and metamyelocytes, and describes their detection in the XT 1800i differential channel by semiconductor laser flow cytometry. It notes a 0.1% detection limit for the analyzer versus a 1% limit for manual counting. This manual is written in Spanish.

What's inside

  • Resumen
  • Abstract
  • Introduccion
  • Material y metodos
  • Resultados
  • Discusion

Frequently asked questions

How many samples with immature granulocyte alarms were analyzed?

120 samples were analyzed.

What correlation coefficient was found between the automated and manual counts?

The correlation coefficient was r = 0.55.

In how many samples were immature granulocytes not detected manually?

20 samples showed no immature granulocytes by manual count.