This study compared the differences between SYSMEX XT1800i automatic classification and manual classification in immature granulocyte alerts. The results showed that the SYSMEX XT1800i automatic classification had a good correlation with the manual classification, but there were still some differences.
This Spanish-language journal article compares the automated leukocyte differential on the Sysmex XT 1800i with a manual differential for immature granulocyte alerts. It reports 120 samples collected over eight weeks, a correlation coefficient of r = 0.55 between absolute values, and concludes that manual counting remains necessary for confirmation.
The article defines immature granulocytes as promyelocytes, myelocytes, and metamyelocytes, and describes their detection in the XT 1800i differential channel by semiconductor laser flow cytometry. It notes a 0.1% detection limit for the analyzer versus a 1% limit for manual counting. This manual is written in Spanish.
120 samples were analyzed.
The correlation coefficient was r = 0.55.
20 samples showed no immature granulocytes by manual count.